Environmental Test Chamber - Climatic Test Chamber – ClimaTT-1000S

Climatic Test Chamber – ClimaTT-1000S

Climatic Test Chamber – ClimaTT-1000S is designed for reliability testing of industrial products. It features precise control of temperature and humidity, along with a broad control range. The performance standards meet the specifications outlined, which details test methods for basic parameters of environmental test equipment used for electric and electronic products, including low temperature, high temperature, constant humidity, and alternating humidity and heat testing equipment.

Sample restrictions

This includes testing and storage of samples containing flammable, explosive, or volatile substances, as well as those involving corrosive materials. It also covers the testing and storage of biological samples, along with the testing or storage of samples that are strong sources of electromagnetic emissions.

Sample requirements

To ensure your test data is accurate and effective, it is essential to follow these guidelines for proper use of the test chamber:

  • The total weight of the load should not exceed 80 kg per cubic meter of the chamber’s internal volume.
  • The total volume of the load should be no more than 1/5 of the chamber’s working space.
  • In any cross-section perpendicular to the main airflow, the load area should not exceed 1/3 of the chamber’s cross-sectional area.
  • Ensure that the placement of the load does not obstruct airflow within the chamber.
  • How Can We Help You?

    Contact our sales team for a quotation.

    Test Conditions
    Equipment cooling method: air-cooled Measured at room temperature + 25℃ with no load. The temperature and humidity performance test is measured according to IEC60068 -3-5 standard. The sensor is placed at the air outlet of the air processing unit.
    Temperature range
    -70℃ ~ +180℃
    Temperature fluctuation
    ±0.5℃
    Temperature uniformity
    ≤2.0℃
    Temperature deviation
    ≤±2.0℃
    Temperature change rate
    Heating rate: -70℃ ~ +100℃, average 3℃/min (no load)
    Cooling rate: +100℃ ~ -70℃, average 1℃/min (no load)
    Maximum load
    N/A
    Humidity range
    20 ~ 98%R.H
    Temperature and humidity range

     

    * With no specimen and under ambient temperature at +25℃.

    * Restrictions on continuous humidity operation at +40℃or lower because of frost on the cooler.

    Humidity deviation
    ±3.0%RH
    Humidity uniformity
    ±3.0%RH
    Humidity fluctuation
    ±2.0%RH
    Noise
    ≤75(dB)(The noise detection device is measured 1maway from the equipment door)
    Test standards
    (MIL-STD.202F) High temperature life test.
    (MIL-STD-810D) High temperature test method.
    (MIL-STD-810D) Low temperature test method.
    (MIL-STD-810D) Damp heat
    Volume
    1000L
    Inner size
    W 1000 mm x H 1000 mm x D 1000 mm
    W 39.37 inches x H 39.37 inches x D 39.37 inches
    Exterior size
    W 1200 mm x H 1925 mm x D 1747 mm ( W 47.24 inches x H 75.78 inches x D 68.77 inches) (Not including external accessories and components) Please refer to the final design confirmation for the external dimensions when ordering.
    Required floor space
    2.07m2 (Should be confirmed at the time of order)